Calibration of a Micromegas-based Gaseous Time Projection Chamber Using Cosmic Ray Muons
arXiv:2305.09297 · doi:10.1088/1748-0221/18/07/P07038
Abstract
We report the calibration of a gaseous Time Projection Chamber based on Micromegas charge readout modules with cosmic ray muons, utilizing their penetrating power and relatively uniform energy deposition per unit length. Muon events were selected through track reconstruction to characterize detector performances, such as the drift velocity, electron lifetime, detector gain, and electric field distortion. The evolution of detector performances with gas purities and electric drift fields over a 50-day data-taking cycle was measured by the muon calibration method. For instance, the drift velocity degraded with gas impurities from to , and then recovered with gas purification. A calibration source was also placed inside the detector as a reference for muon calibrations.
References in corpus (14)
- Dark Matter Search Results from a One TonneYear Exposure of XENON1T
- Dark Matter Search Results from the PandaX-4T Commissioning Run
- Search for Neutrinoless Double-Beta Decay with the Complete EXO-200 Dataset
- Sensitivity of NEXT-100 to neutrinoless double beta decay
- Gaseous and dual-phase time projection chambers for imaging rare processes
- TREX-DM: a low-background Micromegas-based TPC for low-mass WIMP detection
- Characterization of microbulk detectors in argon- and neon-based mixtures
- REST-for-Physics, a ROOT-based framework for event oriented data analysis and combined Monte Carlo response
- Cosmic ray tests of a GEM-based TPC prototype operated in Ar-CF4-isobutane gas mixtures
- High Voltage Insulation and Gas Absorption of Polymers in High Pressure Argon and Xenon Gases
- Design and commissioning of a 600 L Time Projection Chamber with Microbulk Micromegas
- The TPC detector of PandaX-III Neutrinoless Double Beta Decay experiment
- Signal identification with Kalman Filter towards background-free neutrinoless double beta decay searches in gaseous detectors
- Measurement of high-pressure xenon gas absorption in acrylic