Active Personal Eye Lens Dosimetry with the Hybrid Pixelated Dosepix Detector

arXiv:2305.05470

Abstract

Eye lens dosimetry has been an important field of research in the last decade. Dose measurements with a prototype of an active personal eye lens dosemeter based on the Dosepix detector are presented. The personal dose equivalent at mm depth of soft tissue, , was measured in the center front of a water-filled cylinder phantom with a height and diameter of cm. The energy dependence of the normalized response is investigated for mean photon energies between keV and keV for continuous reference radiation fields (N-series) according to ISO 4037. The response normalized to N-60 () at angle of irradiation stays within the approval limits of IEC 61526 for angles of incidence between and . Performance in pulsed photon fields was tested for varying dose rates from up to and pulse durations from up to . The dose measurement works well within the approval limits (acc. to IEC 61526) up to . No significant influence of the pulse duration on the measured dose is found. Reproducibility measurements yield a coefficient of variation which does not exceed for two tested eye lens dosemeter prototypes.

This work has been submitted to the IEEE for possible publication. 9 pages, 10 figures