Variable Energy X-ray Fluorescence Source
arXiv:2211.05691 · doi:10.1088/1748-0221/18/01/P01004
Abstract
We detail the design of a variable energy, x-ray fluorescence source using a low activity (1.8~dpm) \nuc{99}{Tc} source that irradiates thin foils. By rotating the source among foils of Ti, Zn, Nb, Ag, and Au, the device produces x rays between 4 and 70 keV at a rate near 1 Hz. When the source is placed in a storage position, the external radiation is non-detectable. The design of the shielding and rotation mechanism permits use in vacuum and at liquid nitrogen temperature. The design is intended for the study of the low energy response to radiation impinging upon Ge detector surfaces. The source will be useful for understanding the detector response in large-scale Ge arrays such as \textsc{Majorana} and LEGEND.
arXiv commentplanned submission to JINST
References in corpus (7)
- CoGeNT: A Search for Low-Mass Dark Matter using p-type Point Contact Germanium Detectors
- Large-Mass Ultra-Low Noise Germanium Detectors: Performance and Applications in Neutrino and Astroparticle Physics
- Limits on spin-independent couplings of WIMP dark matter with a p-type point-contact germanium detector
- Constraints on Sub-GeV Dark Matter--Electron Scattering from the CDEX-10 Experiment
- First experimental constraints on WIMP couplings in the effective field theory framework from CDEX
- Search for solar axions via axion-photon coupling with the Majorana Demonstrator
- Atomic ionization by sterile-to-active neutrino conversion and constraints on dark matter sterile neutrinos with germanium detectors