Characterization of XIA UltraLo-1800 Response to Measuring Charged Samples
arXiv:2209.08002 · doi:10.1088/1748-0221/18/01/P01027
Abstract
Commercial alpha counters are used in science and industry applications to screen materials for surface radon progeny contamination. In this paper, we characterize an XIA UltraLo-1800, an ionization drift alpha counter, and study the response to embedded charge in polyethylene sample measurements. We show that modeling such effects is possible in a Geant4-based simulation framework and attempt to derive corrections. This paper also demonstrates the effectiveness of the use of an anti-static fan to eliminate the embedded charge and recover a 97.73% alpha detection efficiency in the alpha counter.
References in corpus (5)
- DOLFIN: Automated Finite Element Computing
- Projected Sensitivity of the SuperCDMS SNOLAB experiment
- Construction of arbitrary order finite element degree-of-freedom maps on polygonal and polyhedral cell meshes
- Evaluation and mitigation of trace Pb contamination on copper surfaces
- Radon Daughter Plate-out Measurements at SNOLAB for Polyethylene and Copper