Construction of Continuous Magnetic Cooling Apparatus with Zinc Soldered PrNi Nuclear Stages
arXiv:2112.06490 · doi:10.1007/s10909-022-02801-0
Abstract
We report design details of the whole assembly of a compact and continuous nuclear demagnetization refrigerator (CNDR) with two PrNi nuclear stages, which can keep temperature below 1 mK continuously, and test results of a new thermal contact method for the PrNi stage using Zn soldering rather than Cd soldering. By measuring a residual electrical resistance of a short test piece, the thermal contact resistivity between the PrNi rod and an Ag wires thermal link was estimated as . Based on this value and 2D numerical and 1D analytical thermal simulations, the largest possible temperature gradient throughout the nuclear stage was calculated to be negligibly small ( 2 %) at 1 mK under a 10 nW heat leak, the expected cooling power of the CNDR.
6 pages