The MAJORANA DEMONSTRATOR Readout Electronics System
arXiv:2111.09351 · doi:10.1088/1748-0221/17/05/T05003
Abstract
The MAJORANA DEMONSTRATOR comprises two arrays of high-purity germanium detectors constructed to search for neutrinoless double-beta decay in 76-Ge and other physics beyond the Standard Model. Its readout electronics were designed to have low electronic noise, and radioactive backgrounds were minimized by using low-mass components and low-radioactivity materials near the detectors. This paper provides a description of all components of the MAJORANA DEMONSTRATOR readout electronics, spanning the front-end electronics and internal cabling, back-end electronics, digitizer, and power supplies, along with the grounding scheme. The spectroscopic performance achieved with these readout electronics is also demonstrated.
arXiv commentFor submission to JINST, 17 figures. v2: revised version
References in corpus (18)
- Neutrinoless Double-Beta Decay: Status and Prospects
- Large-Mass Ultra-Low Noise Germanium Detectors: Performance and Applications in Neutrino and Astroparticle Physics
- Search for Zero-Neutrino Double Beta Decay in 76Ge with the Majorana Demonstrator
- A Search for Neutrinoless Double-Beta Decay in Ge with 26 kg-yr of Exposure from the MAJORANA DEMONSTRATOR
- The Majorana Demonstrator Radioassay Program
- New limits on Bosonic Dark Matter, Solar Axions, Pauli Exclusion Principle Violation, and Electron Decay from the Majorana Demonstrator
- First Limit on the Direct Detection of Lightly Ionizing Particles for Electric Charge as Low as /1000 with the \textsc{Majorana Demonstrator}
- Multi-site event discrimination for the MAJORANA DEMONSTRATOR
- Muon Flux Measurements at the Davis Campus of the Sanford Underground Research Facility with the Majorana Demonstrator Veto System
- The Majorana Demonstrator calibration system
- -event Characterization and Rejection in Point-Contact HPGe Detectors
- The Majorana Parts Tracking Database
- ADC Nonlinearity Correction for the MAJORANA DEMONSTRATOR
- Delayed charge recovery discrimination of passivated surface alpha events in P-type point-contact detectors
- High voltage testing for the MAJORANA Demonstrator
- The Sanford Underground Research Facility
- Low Background Materials and Fabrication Techniques for Cables and Connectors in the Majorana Demonstrator
- Contamination Control and Assay Results for the Majorana Demonstrator Ultra Clean Components
Cited by in corpus (10)
- Final Result of the MAJORANA DEMONSTRATOR's Search for Neutrinoless Double- Decay in Ge
- Constraints on the decay of Ta
- Search for solar axions via axion-photon coupling with the Majorana Demonstrator
- Charge Trapping and Energy Performance of the MAJORANA DEMONSTRATOR
- Energy Calibration of Germanium Detectors for the MAJORANA DEMONSTRATOR
- Interpretable Boosted Decision Tree Analysis for the Majorana Demonstrator
- Final Results of the MAJORANA DEMONSTRATOR's Search for Double-Beta Decay of Ge to Excited States of Se
- Rare multi-nucleon decays with the full data sets of the Majorana Demonstrator
- An assay-based background projection for the MAJORANA DEMONSTRATOR using Monte Carlo Uncertainty Propagation
- Machine learning-powered data cleaning for LEGEND: a semi-supervised approach using affinity propagation and support vector machines