Force microscopy cantilevers locally heated in a fluid: temperature fields and effects on the dynamics
arXiv:2110.03404 · doi:10.1063/5.0060911
Abstract
Atomic force microscopy cantilevers are often, intentionally or not, heated at their extremity. We describe a model to compute the resulting temperature field in the cantilever and in the surrounding fluid on a wide temperature range. In air and for common geometries, the heat fluxes in the cantilever and to the environment are of comparable magnitude. We then infer how the fluid-structure interaction is modified due to heating, and predict the induced changes in the dynamics of the system. In particular, we describe how the resonance frequencies of the cantilever shift with a temperature increase due to two competing processes: softening of the cantilever, and decrease of the fluid inertial effects. Our models are illustrated by experiments on a set of cantilevers spanning the relevant geometries to explore the relative importance of both effects.
References in corpus (5)
- Direct Measurement of Room Temperature Non-diffusive Thermal Transport Over Micron Distances in a Silicon Membrane
- Quadrature phase interferometer for high resolution force spectroscopy
- Resonance frequency shift of strongly heated micro-cantilevers
- Silicon cantilevers locally heated from 300K up to the melting point: temperature profile measurement from their resonances frequency shift