Non-galvanic calibration and operation of a quantum dot thermometer
arXiv:2012.01209 · doi:10.1103/PhysRevApplied.15.034044
Abstract
A cryogenic quantum dot thermometer is calibrated and operated using only a single non-galvanic gate connection. The thermometer is probed with radio-frequency reflectometry and calibrated by fitting a physical model to the phase of the reflected radio-frequency signal taken at temperatures across a small range. Thermometry of the source and drain reservoirs of the dot is then performed by fitting the calibrated physical model to new phase data. The thermometer can operate at the transition between thermally broadened and lifetime broadened regimes, and outside the temperatures used in calibration. Electron thermometry was performed at temperatures between and , in both a cryostat and a dilution refrigerator. The experimental setup allows fast electron temperature readout with a sensitivity of , at Kelvin temperatures. The non-galvanic calibration process gives a readout of physical parameters, such as the quantum dot lever arm. The demodulator used for reflectometry readout is readily available and very affordable.
6 pages, 6 figures, supplementary included
References in corpus (7)
- Fast electron thermometry towards ultra-sensitive calorimetric detection
- Primary thermometry triad at 6 mK in mesoscopic circuits
- Multiphoton transitions in Josephson-junction qubits (Review Article)
- Readout methods and devices for Josephson-junction-based solid-state qubits
- Small-signal equivalent circuit for double quantum dots at low-frequencies
- Traceable Coulomb Blockade Thermometry
- Quantum dot thermometry at ultra-low temperature in a dilution refrigerator with a He immersion cell