Calibration of a superconducting gravimeter with an absolute atom gravimeter
arXiv:2007.10291 · doi:10.1007/s00190-021-01516-6
Abstract
We present a 27-days long common view measurement of an absolute cold atom gravimeter (CAG) and a relative iGrav superconducting gravimeter, which we use to calibrate the iGrav scale factor. This allowed us to push the CAG long-term stability down to the level of 0.5~nm.s. We investigate the impact of the duration of the measurement on the uncertainty in the determination of the correlation factor and show that it is limited to about 3\textperthousand~by the coloured noise of our cold atom gravimeter. A 3-days long measurement session with an additional FG5X absolute gravimeter allows us to directly compare the calibration results obtained with two different absolute meters. Based on our analysis, we expect that with an improvement of its long term stability, the CAG will allow to calibrate the iGrav scale factor to better than the per mille level (1 level of confidence) after only one-day of concurrent measurements for maximum tidal amplitudes.
References in corpus (8)
- Gravity measurements below with a transportable absolute quantum gravimeter
- Mobile quantum gravity sensor with unprecedented stability
- Stability comparison of two absolute gravimeters: optical versus atomic interferometers
- Comparison between two mobile absolute gravimeters: optical versus atomic interferometers
- Improving the accuracy of atom interferometers with ultracold sources
- Off-resonant Raman transitions impact in an atom interferometer
- Effective velocity distribution in an atom gravimeter: effect of the convolution with the response of the detection
- Matter-wave laser Interferometric Gravitation Antenna (MIGA): New perspectives for fundamental physics and geosciences