A Portable Brain MRI Scanner for Underserved Settings and Point-Of-Care Imaging
arXiv:2004.13183 · doi:10.1038/s41551-020-00641-5
Abstract
Access to and availability of MRI scanners is typically limited by their cost, siting and infrastructure requirements. This precludes MRI diagnostics, the reference standard for neurological assessment, in patients who cannot be transported to specialized scanner suites. This includes patients who are critically ill and unstable, and patients located in low-resource settings. The scanner design presented here aims to extend the reach of MRI by substantially reducing these limitations. Our goal is to shift the cost-benefit calculation for MRI toward more frequent and varied use, including improved accessibility worldwide and point of care operation. Here, we describe a portable brain MRI scanner using a compact, lightweight permanent magnet, with a built-in readout field gradient. Our low-field (80 mT) Halbach cylinder design of rare-earth permanent magnets results in a 122 kg magnet with minimal stray-field, requiring neither cryogenics nor external power. The built-in magnetic field gradient reduces reliance on high-power gradient drivers, which not only lowers overall system power and cooling requirements, but also reduces acoustic noise. Imperfections in the encoding fields are mitigated with a generalized iterative image reconstruction technique, that uses prior characterization of the field patterns. Our system was validated using T1, T2 and proton density weighted in vivo brain images with a spatial resolution of 2.2 x 1.3 x 6.8 mm.
under review at Nature Biomedical Engineering
Cited by in corpus (13)
- Superconducting magnet designs and MRI accessibility: a review
- Practical Concepts for Design, Construction and Application of Halbach Magnets in Magnetic Resonance
- MaRCoS, an open-source electronic control system for low-field MRI
- High-Performance Permanent Magnet Array Design by a Fast Genetic Algorithm (GA)-based Optimization for Low-Field Portable MRI
- Benchmarking the performance of a low-cost Magnetic Resonance Control System at multiple sites in the open MaRCoS community
- Peripheral Nerve Stimulation limits with fast narrow and broad-band pulses
- Elliptical Halbach magnet and gradient modules for low-field portable MRI
- ezyMRI: How to build an MRI machine from scratch -- Experience from a four-day hackathon
- Portable magnetic resonance imaging of patients indoors, outdoors and at home
- Design and performance of a Toroidal RF Volume Coil with Intrinsic Electromagnetic Interference Rejection for low-field Portable Halbach-Based MRI Systems
- An Exploration of Parallel Imaging System for Very-low Field (50mT) MRI Scanner
- Very-low-field MRI scanners: from the ideal to the real permanent magnet array
- MRI Simulation and Reconstruction Framework for Magnetic Vector Fields