Abstract

Thermodynamic fluctuations in mechanical resonators cause uncertainty in their frequency measurement, fundamentally limiting performance of frequency-based sensors. Recently, integrating nanophotonic motion readout with micro- and nano-mechanical resonators allowed practical chip-scale sensors to routinely operate near this limit in high-bandwidth measurements. However, the exact and general expressions for either thermodynamic frequency measurement uncertainty or efficient, real-time frequency estimators are not well established, particularly for fast and weakly-driven resonators. Here, we derive, and numerically validate, the Cramer-Rao lower bound (CRLB) and an efficient maximum-likelihood estimator for the frequency of a classical linear harmonic oscillator subject to thermodynamic fluctuations. For a fluctuating oscillator without external drive, the frequency Allan deviation calculated from simulated resonator motion data agrees with the derived CRLB for averaging times below, as well as above, the relaxation time . The CRLB approach is general and can be extended to driven resonators, non-negligible motion detection imprecision, as well as backaction from a continuous linear quantum measurement.

11 pages, 2 figures

References in corpus (5)