Fast and Efficient Detection of 511 keV Photons using Cherenkov Light in PbF Crystal coupled to a MCP-PMT and SAMPIC Digitization Module
arXiv:1909.06107 · doi:10.1088/1748-0221/14/12/P12001
Abstract
We study the possibility to use the Cherenkov light for the efficient detection of 511 keV photons with the goal to use it in TOF-PET. We designed and tested two detection modules consisting of PbF crystals attached to Planacon MCP-PMT XP85012. Amplified PMT signals are digitized by the SAMPIC module with high readout rate, up to events/s, and a negligible contribution to the time resolution, below 20 ps (FWHM). We developed a fast 2D scanning system to calibrate the PMT time response and studied in details the timing characteristics of the Planacon PMT. Using a radioactive Na source we measured a detection efficiency of 24% for 511 keV photons in a 10 mm thick crystal and a coincidence resolving time of 280 ps. We analyzed the main factors limiting the time resolution of the large-surface detection module and proposed solutions to improve it, which will be tested in our future project.
arXiv commentAccepted by JINST
References in corpus (7)
- Optical properties of bialkali photocathodes
- Studies on the Cherenkov Effect for Improved Time Resolution of TOF-PET
- Cross-talk suppressed multi-anode MCP-PMT
- Measurements of timing resolution of ultra-fast silicon detectors with the SAMPIC waveform digitizer
- High quantum efficiency S-20 photocathodes in photon counting detectors
- New Approaches for Improvement of TOF-PET
- Studies of MCP-PMTs in the miniTimeCube neutrino detector
Cited by in corpus (4)
- Design study of a scintronic crystal targeting tens of picoseconds time resolution for gamma ray imaging: the ClearMind detector
- Scintillating properties of today available lead tungstate crystals
- High resolution MCP-PMT Readout Using Transmission Lines
- Electro-purification studies and first measurement of relative permittivity of TMBi