Calibration of liquid argon detector with and in different drift field
arXiv:1909.02207
Abstract
and have been used in calibrating a liquid argon (LAr) detector. atoms are produced through the decay of and introduced into the LAr detector through the circulating purification system. The light yield reaches 7.260.02 photonelectrons/keV for 41.5keV from and 7.660.01 photonelectrons/keV for the 511keV from , as a comparison. The light yield varies with the drift electric field from 50 to 200V/cm have been also reported. After stopping fill, the decay rate of with a fitted half-life of 1.830.11 h, which is consistent with the reported value of 1.830.02 h.
arXiv comment6 pages, 11figures, 2 tables