Apparatus for Seebeck coefficient measurement of wire, thin film and bulk materials in the wide temperature range (80-650K)
arXiv:1906.08023 · doi:10.1063/1.5116186
Abstract
A Seebeck coefficient measurement apparatus has been designed and developed, which is very effective for accurate characterization of different type of samples in a wide temperature range (80 - 650K) simultaneously covering low as well as the high-temperature regime. Reducing the complexity of the technical design of sample holder and data collections has always been challenging to implement in a single instrument when samples are in different geometrical shape and electronic structure. Our unique design of sample holder with pressure probes covers measurements of different samples shapes (wires, thin films and pellets) as well as different resistivity ranges (metals, semiconductors and insulators). It is suitable for characterization of different samples sizes (3-12 mm). A double heater configuration powered by a dual channel source meter is employed for maintaining a desired constant temperature difference across the sample for the whole temperature range. Two K-type thermocouples are used for simultaneously reading of temperatures and Seebeck voltages by utilizing different channels of a multichannel digital multimeter. Calibration of the system has been carried out using constantan, chromel and alumel materials and recorded data is found to be very accurate and consistent with earlier reports. The Seebeck coefficients of standard samples of constantan (wire) and GaN (thin film) have been reported, which shows the measurement capability of designed setup with versatile samples.
arXiv comment16 PAGES, 5 FIGURES
References in corpus (4)
- Fabrication of a simple apparatus for the Seebeck coefficient measurement in high temperature region
- Automated Instrumentation for High-Temperature Seebeck Coefficient Measurements
- Fabrication of fully automated setup for high temperature thermal conductivity measurement
- Interfacing of High Temperature Z-meter Setup Using Python
Cited by in corpus (4)
- Thermoelectric Properties of GaN with Carrier Concentration Modulation: An Experimental and Theoretical Investigation
- Wide range temperature-dependent (80-630 K) study of Hall effect and the Seebeck coefficient of \b{eta}-Ga2O3 single crystals
- Instrument for simultaneous measurement of Seebeck coefficient and thermal conductivity in the temperature range 300-800 K with python interfacing
- Efficient and affordable thermoelectric measurement setup using Arduino and LabVIEW for education and research