In this work, we report the fabrication of fully automated experimental setup for high temperature Seebeck coefficient (α) measurement. The K-type thermocouples are used to measure the average temperature of the sample and Seebeck voltage across it. The temperature dependence of the Seebeck coefficients of the thermocouple and its negative leg is taken care by using the integration method. Steady state based differential technique is used for α measurement. Use of limited component and thin heater simplify the sample holder design and minimize the heat loss. The power supplied to the heater decides the temperature difference across the sample and the measurement is carried out by achieving the steady state. The LabVIEW based program is built to automize the whole measurement process. The complete setup is fabricated by using commonly available materials in the market. This instrument is standardized for materials with a wide range of α and for the wide range of ΔT across the specimen. For this purpose, high temperature measurement of α of iron, constantan, bismuth, and Bi_0.36 Sb_1.45 Te_3 samples are carried out and data obtained from these samples are found to be in good agreement with the reported data.
30 pages, 12 figures, Instrumentation Science & Technology (in press)
Cited by in corpus (7)
- Apparatus for Seebeck coefficient measurement of wire, thin film and bulk materials in the wide temperature range (80-650K)
- A novel method for error analysis in radiation thermometry with application to industrial furnaces
- Instrument for simultaneous measurement of Seebeck coefficient and thermal conductivity in the temperature range 300-800 K with python interfacing
- Exploring the possibility of enhancing the figure-of-merit ( 2) of NaCoO: A combined experimental and theoretical study
- Automated Instrumentation for the Determination of the High-Temperature Thermoelectric Figure-of-Merit
- Understanding the Seebeck coefficient of LaNiO3 compound in the temperature range 300-620 K
- Emergence of a hidden-order phase well below the charge density wave transition in a topological Weyl semimetal (TaSe)I