PMT overshoot study for JUNO prototype detector
arXiv:1602.06080 · doi:10.1088/1674-1137/40/9/096002
Abstract
The quality of PMT signal is one of the key items for a large and high precision neutrino experiment, like Daya Bay, JUNO, while most of the experiments are affected by the PMT signal overshoot from its positive HV-single cable scheme. For JUNO prototype detector, we have a detailed study on the PMT overshoot and successfully reduced the ratio of overshoot amplitude to signal to ~1% from previous typical ~10%, with no affection to PMT other parameters. Furthermore, we calculated that the overshoot is a result of discharging of capacitors in the HV-signal splitter and the PMT voltage divider. The study result is extremely important for JUNO and other similar experiments.
6 pages, 11 figures