Characterization of photo-multiplier tubes for the Cryogenic Avalanche Detector
arXiv:1509.02724 · doi:10.1088/1748-0221/10/10/P10010
Abstract
New Cryogenic Avalanche Detector (CRAD) with ultimate sensitivity, that will be able to detect one primary electron released in the cryogenic liquid, is under development in the Laboratory of Cosmology and Particle Physics of the Novosibirsk State University jointly with the Budker Institute of Nuclear Physics. The CRAD will use two sets of cryogenic PMTs in order to get trigger signal either from primary scintillations in liquid Ar or from secondary scintillations in high field gap above the liquid. Two types of cryogenic PMTs produced by Hamamatsu Photonics were tested and the results are presented in this paper. Low background 3 inch PMT R11065- 10 demonstrated excellent performance according to its specifications provided by the producer. The gain measured with single electron response (SER) in liquid Ar reached 10^7, dark count rate rate did not exceed 300 Hz and pulse height resolution of single electron signals was close to 50%(FWHM). However, two R11065-10 PMTs out of 7 tested stopped functioning after several tens minutes of operation immersed completely into liquid Ar. The remaining 5 devices and one R11065-MOD were operated successfully for several hours each with all the parameters according to the producer specifications. Compact 2 inch PMT R6041-506-MOD with metal-channel dynode structure is a candidate for side wall PMT system that will look at electroluminescence in high field region above liquid. Four of these PMTs were tested in liquid Ar and demonstrated gain up to 2x10^7, dark count rate rate below 100 Hz and pulse height resolution of single electron signals of about 110% (FWHM).
Submitted to JINST
References in corpus (2)
Cited by in corpus (11)
- Revealing neutral bremsstrahlung in two-phase argon electroluminescence
- SiPM-matrix readout of two-phase argon detectors using electroluminescence in the visible and near infrared range
- Two-phase Cryogenic Avalanche Detector with electroluminescence gap operated in argon doped with nitrogen
- Measurement of the ionization yield of nuclear recoils in liquid argon using a two-phase detector with electroluminescence gap
- Study of cryogenic photomultiplier tubes for the future two-phase cryogenic avalanche detector
- Puzzling time properties of proportional electroluminescence in two-phase argon detectors for dark matter searches
- Further studies of proportional electroluminescence in two-phase argon
- Observation of unusual slow components in electroluminescence signal of two-phase argon detector
- Study of visible-light emission in pure and methane-doped liquid argon
- Observation of primary scintillations in the visible range in liquid argon doped with methane
- Unraveling the puzzle of slow components in gaseous argon of two-phase detectors for dark matter searches using Thick Gas Electron Multiplier