Response of a proportional counter to Ar and Ge: measured spectra versus Geant4 simulation
arXiv:1509.02669 · doi:10.1016/j.nimb.2016.02.029
Abstract
The energy deposition spectra of Ar and Ge in a miniature proportional counter are measured and compared in detail to the model response simulated with Geant4. A certain modification of the Geant4 code, making it possible to trace the deexcitation of atomic shells properly, is suggested. Modified Geant4 is able to reproduce a response of particle detectors in detail in the keV energy range. This feature is very important for the laboratory experiments that search for massive sterile neutrinos as well as for dark matter searches that employ direct detection of recoil nuclei. This work demonstrates the reliability of Geant4 simulation at low energies.
5 pages, 2 figures; minor changes were done 02 Mar 2016 acording to reviewer's suggestions (English was improved, figures were combined, some words and sentences were corrected including the title), no results and conclusions were changed
References in corpus (4)
- Measurement of the solar neutrino capture rate with gallium metal. III: Results for the 2002--2007 data-taking period
- Light Sterile Neutrinos: A White Paper
- CDMSlite: A Search for Low-Mass WIMPs using Voltage-Assisted Calorimetric Ionization Detection in the SuperCDMS Experiment
- Search for an admixture of sterile neutrino in the electron spectrum from tritium -decay