Long Time Stability of the Energy Scale Calibration of a Quantum 2000
arXiv:1506.03242 · doi:10.1016/j.elspec.2011.08.007
Abstract
According to the international standard ISO 15472 the energy scale of an XPS instrument, type Physical Electronics Quantum 2000, was calibrated. It is shown, how the procedures of the ISO 15472 were adapted to the hardware and software design of the Quantum 2000. The long time stability of the energy scale calibration of the XPS instrument was investigated. The instrumented was operated with a satisfying energy scale calibration over a period of 8 years. All the time energy differences between certain peaks could be measured with the chosen precision of the energy scale calibration.
10 pages, 5 figures, 1 table
Cited by in corpus (4)
- Characterisation of the primary X-ray source of an XPS microprobe Quantum 2000
- Vacuum Quality Monitoring of Analytical Sputter Depth Profile Equipment by Recontamination Measurements of a Sputtered Ti Surface
- Sputter-Induced Cross-Contaminations in Analytical AES and XPS Instrumentation: Utilization of the effect for the In-situ Deposition of Ultrathin Functional Layers
- The Benefit of Wide Energy Range Spectrum Acquisition During Sputter Depth Profile Measurements