Evaluation of new large area PMT with high quantum efficiency
arXiv:1504.03174 · doi:10.1088/1674-1137/40/2/026002
Abstract
The neutrino detector of the Jiangmen Underground Neutrino Observatory (JUNO) is designed to use 20 kilotons of liquid scintillator and approximately 16,000 20-inch photomultipliers (PMTs).One of the options is to use the 20-inch R12860 PMT with high quantum efficiency which has recently been developed by Hamamatsu Photonics. The performance of the newly developed PMT preproduction samples is evaluated. The results show that its quantum efficiency is at . Its Peak/Valley (P/V) ratio for the single photoelectron is 4.75 and the dark count rate is at the threshold of 3 mV while the gain is at . The transit time spread of a single photoelectron is . Generally the performances of this new 20-inch PMT are improved over the old one of R3600.
arXiv comment7 pages,12 figures,submitted to Chinese Physics C
References in corpus (2)
Cited by in corpus (7)
- Mass Testing and Characterization of 20-inch PMTs for JUNO
- Gain and charge response of 20 inch MCP and dynode PMTs
- Measurement of the relative Quantum Efficiency of Hamamatsu model R5912-20MOD photomultiplier tubes at liquid argon temperature
- Study of 20-inch PMTs dark count generated large pulses
- A study of the new hemispherical 9-inch PMT
- Characterization of the Hamamatsu 8-inch R14688-100 PMT
- An Alternative Design for Large Scale Liquid Scintillator Detectors