The Photomultiplier Tube Calibration System of the MicroBooNE Experiment
arXiv:1502.04159 · doi:10.1088/1748-0221/10/06/T06001
Abstract
We report on the design and construction of a LED-based fiber calibration system for large liquid argon time projection detectors. This system was developed to calibrate the optical systems of the MicroBooNE experiment. As well as detailing the materials and installation procedure, we provide technical drawings and specifications so that the system may be easily replicated in future LArTPC detectors.