Determining the 6Li Doped Side of a Glass Scintillator for Ultra Cold Neutrons
arXiv:1502.01392 · doi:10.1016/j.nima.2015.04.022
Abstract
Ultracold neutron (UCN) detectors using two visually very similar, to the microscopic level, pieces of optically contacted Cerium doped lithium glasses have been proposed for high rate UCN experiments. The chief difference between the two glass scintillators is that one side is 6Li depleted and the other side Li doped. This note outlines a method to determine which side of the glass stack is doped with 6Li using AmBe and 252Cf neutron sources, and a Si surface barrier detector. The method sees an excess of events around the alpha and triton energies of neutron capture on 6Li when the enriched side is facing the Si surface barrier detector.
5 pages, 5 figures, submitted to Nucl. Instr. and Meth. A
References in corpus (2)
Cited by in corpus (3)
- Ultracold neutron detection with 6Li-doped glass scintillators, NANOSC: a fast ultracold neutron detector for the nEDM experiment at the Paul Scherrer Institute
- Characterization of a scintillating lithium glass ultra-cold neutron detector
- Response of a Li-glass/multi-anode photomultiplier detector to -particles from Am