Maximum-likelihood fits to histograms for improved parameter estimation
arXiv:1312.5622 · doi:10.1007/s10909-014-1098-4
Abstract
Straightforward methods for adapting the familiar chi^2 statistic to histograms of discrete events and other Poisson distributed data generally yield biased estimates of the parameters of a model. The bias can be important even when the total number of events is large. For the case of estimating a microcalorimeter's energy resolution at 6 keV from the observed shape of the Mn K-alpha fluorescence spectrum, a poor choice of chi^2 can lead to biases of at least 10% in the estimated resolution when up to thousands of photons are observed. The best remedy is a Poisson maximum-likelihood fit, through a simple modification of the standard Levenberg-Marquardt algorithm for chi^2 minimization. Where the modification is not possible, another approach allows iterative approximation of the maximum-likelihood fit.
Accepted by J. Low Temperature Physics, in the Proceedings of Low Temperature Detectors 15
Cited by in corpus (7)
- A Reassessment of Absolute Energies of the X-ray L Lines of Lanthanide Metals
- Microcalorimeter Spectroscopy at High Pulse Rates: a Multi-Pulse Fitting Technique
- Absolute energies and emission line shapes of the L x-ray transitions of lanthanide metals
- When "Optimal Filtering" Isn't
- On low-energy tail distortions in the detector responsefunction of x-ray microcalorimeter spectrometers
- System performance of a cryogenic test-bed for the time-division multiplexing readout for NewAthena X-IFU
- The first cut is the cheapest: optimizing Athena/X-IFU-like TES detectors resolution by filter truncation