Studies on the Cherenkov Effect for Improved Time Resolution of TOF-PET
arXiv:1305.6257 · doi:10.1109/TNS.2013.2281667
Abstract
With the newly gained interest in the time of flight method for positron emission tomography (TOF-PET), many options for pushing the time resolution to its borders have been investigated. As one of these options the exploitation of the Cherenkov effect has been proposed, since it allows to bypass the scintillation process and therefore provides almost instantaneous response to incident 511keV annihilation photons. Our simulation studies on the yield of Cherenkov photons, their arrival rate at the photon detector and their angular distribution reveal a significant influence by Cherenkov photons on the rise time of inorganic scintillators - a key-parameter for TOF in PET. A measurement shows the feasibility to detect Cherenkov photons in this low energy range.
References in corpus (1)
Cited by in corpus (7)
- Dosimetry for FLASH Radiotherapy: A Review of Tools and the Role of Radioluminescence and Cherenkov Emission
- Ultrafast nanocomposite scintillators based on Cd-enhanced CsPbCl3 nanocrystals in polymer matrix
- Fast and Efficient Detection of 511 keV Photons using Cherenkov Light in PbF Crystal coupled to a MCP-PMT and SAMPIC Digitization Module
- Monte Carlo study of the Coincidence Resolving Time of a liquid xenon PET scanner, using Cherenkov radiation
- Precision timing detectors
- Rethinking Timing Residuals: Advancing PET Detectors with Explicit TOF Corrections
- A High Count-Rate and Depth-of-Interaction Resolving Single Layered One-Side Readout Pixelated Scintillator Crystal Array for PET Applications