Micromegas-TPC operation at high pressure in xenon-trimethylamine mixtures
arXiv:1210.3287 · doi:10.1088/1748-0221/8/01/P01012
Abstract
In this work we present a systematic study of Micromegas detectors in high pressure gaseous Xenon using trimethylamine (TMA) as quencher gas. Gas gains and energy resolutions for 22.1 keV X-rays are measured for pressures between 1 and 10 bar and various relative concentrations of TMA from 0.3 % to 15 %. We observe stable operation at all pressures, and a strongly enhanced gas gain, suggestive of Penning-like energy-transfer processes. The effect is present at all pressures and it is strongest at TMA concentrations ranging from 1.5 % to 3 %. Operating in this concentration range, the maximum gain reached values as high as x10^3 (x10^2) at 1 (10) bar. Besides, the energy resolution achievable for 22.1 keV X-rays is substantially better than the one previously obtained in pure Xe, going down to 7.3 % (9.6 %) FWHM for 1 (10) bar. These results are of interest for calorimetric applications of high pressure gas Xe TPCs, in particular for the search of the neutrinoless double beta decay of Xe-136. The resolutions achieved would extrapolate into 0.7 % (0.9 %) FWHM at the Qbb value of Xe-136 for 1 (10) bar.
References in corpus (7)
- Search for Neutrinoless Double-Beta Decay in Xe with EXO-200
- NEXT-100 Technical Design Report (TDR). Executive Summary
- Radiopurity of Micromegas readout planes
- Characterization of microbulk detectors in argon- and neon-based mixtures
- Micromegas operation in high pressure xenon: charge and scintillation readout
- Micromegas readouts for double beta decay searches
- Status of R&D on Micromegas for Rare Event Searches: The T-REX project
Cited by in corpus (25)
- Readout technologies for directional WIMP Dark Matter detection
- Gaseous and dual-phase time projection chambers for imaging rare processes
- Accurate gamma and MeV-electron track reconstruction with an ultra-low diffusion Xenon/TMA TPC at 10 atmospheres
- Gaseous time projection chambers for rare event detection: Results from the T-REX project. I. Double beta decay
- Background assessment for the TREX Dark Matter experiment
- Characterization of a medium size Xe/TMA TPC instrumented with microbulk Micromegas, using low-energy -rays
- Neutrino-hydrogen interactions with a high-pressure time projection chamber
- Description and commissioning of NEXT-MM prototype: first results from operation in a Xenon-Trimethylamine gas mixture
- Exploring 0.1-10eV axions with a new helioscope concept
- Dielectric Strength of Noble and Quenched Gases for High Pressure Time Projection Chambers
- Modelling the behaviour of microbulk Micromegas in Xenon/trimethylamine gas
- Design and commissioning of a 600 L Time Projection Chamber with Microbulk Micromegas
- Directional dark matter by polar angle direct detection and application of columnar recombination
- Microbulk Micromegas for the search of 0 of 136Xe in the PandaX-III experiment
- Study of the loss of Xenon Scintillation in Xenon-Trimethylamine Mixtures
- A new amplification structure for time projection chambers based on electroluminescence
- Micromegas-TPC operation at high pressure in Xenon-trimethylamine mixtures
- Detectors and Concepts for sub-100 ps timing with gaseous detectors
- Lessons from the operation of the "Penning-Fluorescent" TPC and prospects
- TREX-DM: a low background Micromegas-based TPC for low mass WIMP detection
- Pressure effect in the X-ray intrinsic position resolution in noble gases and mixtures
- Results from the NEXT prototypes
- TREX-DM: a low background Micromegas-based TPC for low-mass WIMP detection
- Enhanced search sensitivity to the double beta decay of Xe to excited states with topological signatures
- Characterization of a Spherical Proportional Counter in argon-based mixtures