Design of an optical reference cavity with low thermal noise and flexible thermal expansion properties
arXiv:1204.1209 · doi:10.1140/epjd/e2013-30458-2
Abstract
An ultrastable optical reference cavity with re-entrant fused silica mirrors and a ULE spacer structure is designed through finite element analysis. The designed cavity has a low thermal noise limit of and a flexible zero crossing temperature of the effective coefficient of thermal expansion (CTE). The CTE zero crossing temperature difference between a composite cavity and a pure ULE cavity can be tuned from C to C, which enables operation of the designed reference cavity near room temperature without worrying about the CTE zero crossing temperature of the ULE spacer. The design can be applied to cavities with different lengths. Vibration immunity of the cavity is also achieved through structure optimization.
arXiv comment16 pages, 12 figures
References in corpus (15)
- A sub-40 mHz linewidth laser based on a silicon single-crystal optical cavity
- Sr lattice clock at 1x10^{-16} fractional uncertainty by remote optical evaluation with a Ca clock
- Making optical atomic clocks more stable with level laser stabilization
- Compact, thermal-noise-limited optical cavity for diode laser stabilization at 1 x 10-15
- New Limits on Coupling of Fundamental Constants to Gravity Using Sr Optical Lattice Clocks
- Sub-Hz line width diode lasers by stabilization to vibrationally and thermally compensated ULE Fabry-Perot cavities
- Ultrastable lasers based on vibration insensitive cavities
- Observation of the 1S0 - 3P0 clock transition in 27Al+
- Tuning the Thermal Expansion Properties of Optical Reference Cavities with Fused Silica Mirrors
- An accurate optical lattice clock with 87Sr atoms
- Sub-hertz frequency stabilization of a commercial diode laser
- A vibration-insensitive optical cavity and absolute determination of its ultrahigh stability
- Some aspects of simulation and realization of an optical reference cavity
- Spin-1/2 Optical Lattice Clock
- Towards an ultra-stable optical sapphire cavity system for testing Lorentz invariance