A single-solenoid pulsed-magnet system for single-crystal scattering studies
arXiv:1109.6713 · doi:10.1063/1.3688251
Abstract
We present a pulsed-magnet system that enables x-ray single-crystal diffraction in addition to powder and spectroscopic studies with the magnetic field applied on or close to the scattering plane. The apparatus consists of a single large-bore solenoid, cooled by liquid nitrogen. A second independent closed-cycle cryostat is used for cooling samples near liquid helium temperatures. Pulsed magnetic fields close to T with a zero-to-peak-field rise time of 2.9 ms are generated by discharging a 40 kJ capacitor bank into the magnet coil. The unique characteristic of this instrument is the preservation of maximum scattering angle () on the entrance and exit sides of the magnet bore by virtue of a novel double-funnel insert. This instrument will facilitate x-ray diffraction and spectroscopic studies that are impractical, if not impossible, to perform using split-pair and narrow-opening solenoid magnets, and offers a practical solution for preserving optical access in future higher-field pulsed magnets.
Accepted for publication in Rev. Sci. Instrum
References in corpus (5)
- Charge order, dynamics, and magneto-structural transition in multiferroic LuFeO
- Magnetoelastics of a spin liquid: X-ray diffraction studies of Tb2Ti2O7 in pulsed magnetic fields
- Direct observation of the high magnetic field effect on the Jahn-Teller state in TbVO4
- Time-resolved one-dimensional detection of x-ray scattering in pulsed magnetic fields
- A robust but disordered collapsed-volume phase in a cerium alloy under the application of pulsed magnetic fields
Cited by in corpus (6)
- Structural evolution of the kagome superconductors VSb ( = K, Rb, and Cs) through charge density wave order
- The Tricritical Point of the f-electron Antiferromagnet USb2 Driven by High Magnetic Fields
- Search for Two-Photon Interaction with Axionlike Particles Using High-Repetition Pulsed Magnets and Synchrotron X Rays
- A 31T split-pair pulsed magnet for single crystal x-ray diffraction at low temperature
- Measurement of the angle dependence of magnetostriction in pulsed magnetic fields using a piezoelectric strain gauge
- Single shot x-ray diffractometry in SACLA with pulsed magnetic fields up to 16 T