Comparison of discharge lamp and laser pumped cesium magnetometers
arXiv:physics/0412011 · doi:10.1007/s00340-005-1773-x
Abstract
We have performed a comparison of laser (LsOPM) and lamp (LpOPM) pumped cesium vapor magnetometers. Although the LsOPM operated 50% above its shot-noise limit we found an intrinsic sensitivity of 15 fT/\sqrt{Hz} and 25 fT/\sqrt{Hz} for the LsOPM and LpOPM respectively. Two modes of operation, viz.,the phase-stabilized and the self-oscillating mode were investigated and found to yield a similar performance. We have compared the performance of the LsOPM and the LpOPM directly by simultaneous measurements of field fluctuations of a 2 microtesla magnetic field inside a multilayer magnetic shield and have used one of the magnetometers for an active field stabilization. In the stabilized mode we found a gradient instability of 25 fT within an integration time of 100 s, which represents an upper limit of the long-term stability of the magnetometers.
11 pages, 13 figures, 1 table
References in corpus (1)
Cited by in corpus (12)
- Optical Magnetometry
- Flavour physics of leptons and dipole moments
- A Low-Noise High-Density Alkali Metal Scalar Magnetometer
- Atomic magnetic resonance induced by amplitude-, frequency-, or polarization-modulated light
- Test of Lorentz invariance with spin precession of ultracold neutrons
- Synchronous Optical Pumping of Quantum Revival Beats for Atomic Magnetometery
- Experimental study of laser detected magnetic resonance based on atomic alignment
- Heading-error-free optical atomic magnetometry in the Earth-field range
- A sound card based multi-channel frequency measurement system
- Optically Pumped Magnetometer with High Spatial Resolution Magnetic Guide for the Detection of Magnetic Droplets in a Microfluidic Channel
- Experimental study of 199Hg spin anti-relaxation coatings
- Neutron Electric Dipole Moment Experiments