Measurements and simulations of Cherenkov light in lead fluoride crystals
arXiv:nucl-ex/0108019 · doi:10.1016/S0168-9002(01)00668-4
Abstract
The anticipated use of more than one thousand lead fluoride (PbF2) crystals as a fast and compact Cherenkov calorimeter material in a parity violation experiment at MAMI stimulated the investigation of the light yield (L.Y.) of these crystals. The number of photoelectrons (p.e.) per MeV deposited energy has been determined with a hybrid photomultiplier tube (HPMT). In response to radioactive sources a L.Y. between 1.7 and 1.9 p.e./MeV was measured with 4% statistical and 5% systematic error. The L.Y. optimization with appropriate wrappings and couplings was investigated by means of the HPMT. Furthermore, a fast Monte Carlo simulation based on the GEANT code was employed to calculate the characteristics of Cherenkov light in the PbF2 crystals. The computing time was reduced by a factor of 50 compared to the regular photon tracking method by implementing detection probabilities as a three-dimensional look-up table. For a single crystal a L.Y. of 2.1 p.e./MeV was calculated. The corresponding detector response to electrons between 10 and 1000 MeV was highly linear with a variation smaller than 1%.
Cited by in corpus (9)
- Measurement of Strange Quark Contributions to the Nucleon's Form Factors at Q^2=0.230 (GeV/c)^2
- Measurement of Strange Quark Contributions to the Vector Form Factors of the Proton at Q**2=0.22 (GeV/c)**2
- Studies of an array of PbF2 Cherenkov crystals with large-area SiPM readout
- The P2 Experiment - A future high-precision measurement of the electroweak mixing angle at low momentum transfer
- Realtime calibration of the A4 electromagnetic lead fluoride calorimeter
- A high power liquid hydrogen target for the Mainz A4 parity violation experiment
- A luminosity monitor for the A4 parity violation experiment at MAMI
- Design and Performance of a Lead Fluoride Detector as a Luminosity Monitor
- Lead fluoride Cherenkov detector read out by avalanche photodiodes for measuring the intensities of pulsed antiproton beams