We describe the design, construction and performance of the PIBETA detector built for the precise measurement of the branching ratio of pion beta decay, pi+ -> pi0 e+ nu, at the Paul Scherrer Institute. The central part of the detector is a 240-module spherical pure CsI calorimeter covering 3*pi sr solid angle. The calorimeter is supplemented with an active collimator/beam degrader system, an active segmented plastic target, a pair of low-mass cylindrical wire chambers and a 20-element cylindrical plastic scintillator hodoscope. The whole detector system is housed inside a temperature-controlled lead brick enclosure which in turn is lined with cosmic muon plastic veto counters. Commissioning and calibration data were taken during two three-month beam periods in 1999/2000 with pi+ stopping rates between 1.3*E3 pi+/s and 1.3*E6 pi+/s. We examine the timing, energy and angular detector resolution for photons, positrons and protons in the energy range of 5-150 MeV, as well as the response of the detector to cosmic muons. We illustrate the detector signatures for the assorted rare pion and muon decays and their associated backgrounds.
117 pages, 48 Postscript figures, 5 tables, Elsevier LaTeX, submitted to Nucl. Instrum. Meth. A
References in corpus (6)
- Cosmic muon tomography of pure cesium iodide calorimeter crystals
- Radiative corrections of $O(\a)$ for pion beta decay in the light-front quark model
- The current status of Vud
- Light response of pure CsI calorimeter crystals painted with wavelength-shifting lacquer
- ADC common noise correction and zero suppression in the PIBETA detector
- Measurement of stopping beam distributions in the PIBETA detector
Cited by in corpus (18)
- New physics searches in nuclear and neutron decay
- Precise Measurement of the Pi+ -> Pi0 e+ nu Branching Ratio
- Detailed Report of the MuLan Measurement of the Positive Muon Lifetime and Determination of the Fermi Constant
- New Precise Measurement of the Pion Weak Form Factors in the Pi+ -> e+ nu gamma Decay
- Precise Measurement of the Pion Axial Form Factor in the Pi+ -> e+ nu gamma Decay
- Constraining axion-like particles from rare pion decays
- Comparison between high-energy proton and charged pion induced damage in Lead Tungstate calorimeter crystals
- Experimental study of rare charged pion decays
- An apparatus for studying spallation neutrons in the Aberdeen Tunnel laboratory
- New results in rare allowed muon and pion decays
- Energetic neutron identification with pulse shape discrimination in pure CsI crystals
- Data acquisition system for the MuLan muon lifetime experiment
- Probing the Standard Model via rare pion and muon decays
- Beam particle tracking with a low-mass mini time projection chamber in the PEN experiment
- Precise Measurement of pi+ -> e+ nu Branching Ratio
- Measurement of stopping beam distributions in the PIBETA detector
- The automatic gain-matching in the PIBETA CsI calorimeter
- PEN: a low energy test of lepton universality