Laboratory Characterization of a Ge-CCD Prototype for Future Near-Infrared Astronomical Instrumentation

arXiv:2610.22785

Abstract

Germanium charge-coupled devices (Ge-CCDs) extend CCD imaging beyond the silicon cutoff and are candidate detectors for the near-infrared channel of future astronomical instruments, in particular the spectrographs of next-generation wide-field surveys. We report the laboratory characterization at Brookhaven National Laboratory (BNL) of a prototype fabricated at MIT Lincoln Laboratory. We describe the cryogenic test system and readout, and present measurements of the output waveform, dark current, readout noise, and X-ray response. The dark current at 94.69 K is e pixel s. Under Fe irradiation without on-chip binning, the charge generated by an X-ray is confined within pixels, and the Mn K line has a full width at half maximum (FWHM) of % for single-pixel events. Since the charge generated by X-rays above about 30 keV spreads over or more pixels, the CCD was operated with on-chip binning at higher energies. This yielded FWHMs of % at 30.9 keV (Ba, ASCA grades 0, 2, 3, 4, and 6) and % at 59.5 keV (Am, all grades). These results establish a baseline for the device and for the BNL test system. Measurements of the temperature dependence, optical response, and charge-transfer performance will connect this baseline to the requirements of near-infrared astronomical instruments.