A Portable Dual-Color Two-Photon Rubidium Optical Frequency Standard
arXiv:2609.08083
Abstract
Portable atomic clocks are essential in a wide variety of applications, most notably in the operation of global navigation satellite systems. Existing portable atomic clocks utilizing microwave-based interrogation schemes are now routinely eclipsed by the next generation of atomic frequency standards based on optical interrogation. While optical frequency standards demonstrate greatly improved frequency stability, they have only recently reached a level of technical maturity required to demonstrate this improved performance outside of well curated laboratory environments. Here, we demonstrate a fully autonomous and portable optical frequency standard based on an efficient dual-color excitation of the two-photon transition in Rb. The standard utilizes a combination of robust, highly developed commercial-off-the-shelf telecommunications technologies and a fully integrated portable optical frequency comb, providing the optical and microwave outputs vital for interfacing with existing electronic systems and infrastructure. The system demonstrates a fractional frequency stability of at 1s of integration time, reaching at 8000s of integration time without the need for drift removal. This portable demonstrator unit marks a significant achievement in the development of Rb optical atomic frequency standards, and for the deployment of optical atomic frequency standards outside of the laboratory.