Resolution-Noise Characteristics of Common FDK Filter Kernels: A Practical Reference for Preclinical Cone-Beam Micro-CT
arXiv:2605.00239 · doi:10.1371/journal.pone.0349360
Abstract
The ramp filter kernel and cutoff frequency are fundamental parameters of the Feldkamp-Davis-Kress (FDK) algorithm that determine the resolution and noise characteristics of the reconstructed image. Despite their importance, systematic evaluations of their combined effect on task-based image quality in preclinical micro-CT are scarce, and many studies do not report the filter configuration used. We reconstruct identical data from a GE eXplore CT 120 scanner using four filter kernels (ramp, Shepp-Logan, cosine, Hamming) at four cutoff frequencies (1.0, 0.8, 0.6, and Nyquist, matched to the detector-to-voxel size ratio) and evaluate each of the sixteen configurations using the modulation transfer function (MTF), noise power spectrum (NPS), and non-prewhitening detectability index (NPW ). Qualitative assessment is performed on a mouse lung specimen. Across the sixteen configurations, ranges from 0.93 to 2.35 lp/mm, integrated NPS from 75,670 to 13,259 , and the Rose criterion crossing diameter from 2.86 to 0.93 mm at HU and from 7.74 to 3.62 mm at 100 HU. This note presents the data as a concise visual and quantitative reference for groups selecting FDK filter parameters for preclinical cone-beam CT.