A Top-Loading Point-Contact Spectroscopy Probe with In-Situ Sample Exchange for Dilution Refrigerators
arXiv:2604.03821
Abstract
We report the design and implementation of a point-contact spectroscopy (PCS) system integrated with a dilution refrigerator, enabling measurements down to 30 mK. The setup employs a needle-anvil geometry with a cryogenic piezo-driven nanopositioner for in-situ formation of mesoscopic point contacts. We discuss the thermal anchoring strategies that enable efficient cooling of the probe to ultra-low temperatures and reliable measurements. We also address positioner-related challenges and the solutions implemented to ensure stable operation at millikelvin temperatures. The performance of the probe is demonstrated through point contact spectroscopy on Ta-doped TiSe (TaTiSe, ), a superconductor with K. The spectra exhibit well-defined superconducting features that systematically diminish with increasing temperature and magnetic field. The platform provides a robust and versatile tool for spectroscopic investigations of superconductors and other quantum materials at millikelvin temperatures and high magnetic fields.
9pages ,6 figures