Performance Characterization of a Plastic-Scintillator Sensor for Fast-Neutron, Thermal-Neutron, and Gamma-Ray Discrimination
arXiv:2512.15585 · doi:10.3390/s26113462
Abstract
Discrimination of fast neutrons, thermal neutrons, and rays in mixed radiation fields is important for radiation monitoring, reactor-related measurements, and background suppression in nuclear experiments. In this work, we investigate a compact plastic-scintillator sensor composed of EJ276 or EJ200 optically coupled to an EJ426 thermal neutron screen and read out by a single photomultiplier tube (PMT). The equivalent energy response of the detector assemblies was calibrated using Cs, Na, and Co sources through Compton edge analysis, and pulse shape discrimination was evaluated with an AmBe neutron source under different moderator thicknesses. The EJ200+EJ426 assembly provides a well separated discrimination between thermal-neutron capture events and dominated events over the measured range, with a figure of merit greater than 5. In contrast, the EJ276+EJ426 assembly produced three identifiable signal populations associated with rays, fast neutrons, and thermal neutrons. These results show that the proposed sensor architecture is a promising compact approach for mixed field radiation applications