An Open-Path Eddy-Covariance Laser Spectrometer for Simultaneous Monitoring of CO, CH, and HO
arXiv:2512.01619 · doi:10.3390/s26020462
Abstract
We present E-CAHORS a compact mid-infrared open-path diode-laser spectrometer designed for the simultaneous measurement of carbon dioxide, methane, and water vapor concentrations in the near-surface atmospheric layer. These measurements, combined with simultaneous data from a three-dimensional anemometer, can be used to determine fluxes using the eddy-covariance method. The instrument utilizes two interband cascade lasers operating at 2.78 m and 3.24 m within a novel four-pass M-shaped optical cell, which provides high signal power and long-term field operation without requiring active air sampling. Two detection techniques tunable diode laser absorption spectroscopy (TDLAS) and a simplified wavelength modulation spectroscopy (sWMS) were implemented and evaluated. Laboratory calibration demonstrated linear responses for all gases (R 0.999) and detection precisions at 10 Hz of 311 ppb for CO, 8.87 ppb for CH, and 788 ppb for HO. Field tests conducted at a grassland site near Moscow showed strong correlations (R = 0.91 for CO and HO, R = 0.67 for CH) with commercial LI-COR LI-7200 and LI-7700 analyzers. The TDLAS mode demonstrated lower noise and greater stability under outdoor conditions, while sWMS provided baseline-free spectra but was more sensitive to power fluctuations. E-CAHORS combines high precision, multi-species sensing capability with low power consumption (10 W) and a compact design (4.2 kg).