paper

A Positivity-Preserving Finite Element Framework for Accurate Dose Computation in Proton Therapy

arXiv:2506.01105

Abstract

We present a stabilised finite element method for modelling proton transport in tissue, incorporating both inelastic energy loss and elastic angular scattering. A key innovation is a positivity-preserving formulation that guarantees non-negative fluence and dose, even on coarse meshes. This enables reliable computation of clinically relevant quantities for treatment planning. We derive a priori error estimates demonstrating optimal convergence rates and validate the method through numerical benchmarks. The proposed framework provides a robust, accurate and efficient tool for advancing proton beam therapy.

21 pages, 11 figures

A Positivity-Preserving Finite Element Framework for Accurate Dose Computation in Proton Therapy · wovepaper