paper

Endoscopic fiber-coupled diamond magnetometer for cancer surgery

arXiv:2504.05884 · doi:10.1103/znt3-988w

Abstract

Interoperative measurements using magnetic sensors is a valuable technique in cancer surgery for finding magnetic tracers. Here we present a fiber-coupled nitrogen-vacancy (N-V) center magnetometer capable of detecting iron oxide suspension (MagTrace from Endomagnetics Ltd.) used in breast cancer surgeries. Detection of an iron mass as low as 0.56~mg has been demonstrated, 100 times less than that of a recommended dose at a maximum distance of 5.8~mm. Detection of an iron concentration as low as 2.8 mg/ml has also been demonstrated, 20 times less than a recommended dose. The maximum working distance from the sensor can be as large as 14.6~mm for higher concentrations. The sensor head has a maximum diameter of 10~mm which would allow it to be used for endoscopy, laparoscopy and interoperative surgery.

9 pages, 15 figures

References in corpus (1)