Comparison of the detector response and calibration function of metallic microcalorimeters for X-ray photons and external electrons
arXiv:2502.05975 · doi:10.1016/j.nima.2025.170662
Abstract
Metallic microcalorimeters (MMCs) are cryogenic single-particle detectors that rely on a calorimetric detection principle. Due to their excellent energy resolution, close-to-ideal linear detector response, fast signal rise time and the potential for \SI{100}{\%} quantum efficiency, MMCs outperform conventional detectors by several orders of magnitude in resolution. These attributes make them particularly interesting for a broad spectrum of applications, including a next-generation neutrino mass experiment based on the measurement of the tritium beta-decay spectrum, with an objective of achieving a sensitivity surpassing that of the pioneering KATRIN experiment. However, although MMCs have been used in measurements of photons and heavy ions with great success, no information is currently available on the interaction between MMCs and external light charged particles such as electrons. This work aims to provide such missing information and to demonstrate that MMC-based detectors are suitable for high-resolution spectroscopy of external electron sources. Particularly, we present the first-ever measurements of external electrons using a metallic microcalorimeter, comprehensively discuss the characteristics of the signal shape and the calibration function and give a direct comparison between well-defined conversion electron and X-ray photon signals from the same Rb/Kr source.
18 pages, 8 figures. Nuclear Instruments and Methods in Physics Research - section A (NIM-A)
References in corpus (9)
- Evidence for oscillation of atmospheric neutrinos
- Measurement of the rate of nu_e + d --> p + p + e^- interactions produced by 8B solar neutrinos at the Sudbury Neutrino Observatory
- Neutrino cosmology and Planck
- Direct neutrino-mass measurement based on 259 days of KATRIN data
- Direct Measurements of Neutrino Mass
- -Decay Spectrum, Response Function and Statistical Model for Neutrino Mass Measurements with the KATRIN Experiment
- Magnetic microcalorimeter with paramagnetic temperature sensors and integrated dc-SQUID readout for high-resolution X-ray emission spectroscopy
- Metallic magnetic calorimeter arrays for the first phase of the ECHo experiment
- Optimum filter-based analysis for the characterization of a high-resolution magnetic microcalorimeter towards the DELight experiment