We demonstrate a novel shot-to-shot acquisition method for optical pump - keV electron energy probe in ultrafast scattering experiments. We integrate a phase-locked acquisition scheme at a repetition rate of 20kHz in a conventional ultrafast electron diffraction (UED) setup. We proceed to a full characterization of the noise level in different configurations and for realistic scenarios. The shot-to-shot acquisition improves the signal-to-noise ratio (SNR) by one order of magnitude and can be readily implemented in other high-repetition rate electron diffraction and spectroscopy setups.
Submitted in Structural Dynamics
References in corpus (14)
- Polarized phonons carry the missing angular momentum in femtosecond demagnetization
- Laser-Induced Skyrmion Writing and Erasing in an Ultrafast Cryo-Lorentz Transmission Electron Microscopy
- Coherent and incoherent electron-phonon coupling in graphite observed with radio-frequency compressed ultrafast electron diffraction
- Mapping momentum-dependent electron-phonon coupling and non-equilibrium phonon dynamics with ultrafast electron diffuse scattering
- Momentum-Resolved View of Electron-Phonon Coupling in Multilayer WSe
- Direct determination of mode-projected electron-phonon coupling in the time-domain
- Compact femtosecond electron diffractometer with 100 keV electron bunches approaching the single-electron pulse duration limit
- Nanoscale mapping of ultrafast magnetization dynamics with femtosecond Lorentz microscopy
- Solving the Jitter Problem in Microwave Compressed Ultrafast Electron Diffraction Instruments: Robust Sub-50 fs Cavity-Laser Phase Stabilization
- Time- and momentum-resolved phonon population dynamics with ultrafast electron diffuse scattering
- Does VO Host a Transient Monoclinic Metallic Phase?
- Ultrafast formation of topological defects in a 2D charge density wave
- Imaging the ultrafast coherent control of a skyrmion crystal
- Magnon diffuse scattering in scanning transmission electron microscopy