paper

Calibration of Scanning Thermal Microscope using Optimal Estimation of Function Parameters by Iterated Linearization

arXiv:2501.08961

Abstract

Scanning thermal microscopy is a unique tool for the study of thermal properties at the nanoscale. However, calibration of the method is a crucial problem. When analyzing local thermal conductivity, direct calibration is not possible and reference samples are used instead. As the calibration dependence is non-linear and there are only a few calibration points, this represents a metrological challenge that needs complex data processing. In this contribution we present use of the OEFPIL algorithm for robust and single-step evaluation of local thermal conductivities and their uncertainties, simplifying this procedure. Furthermore, we test the suitability of SThM calibration for automated measurement.

Calibration of Scanning Thermal Microscope using Optimal Estimation of Function Parameters by Iterated Linearization · wovepaper