Determination of the force transmission error in a single-sinker magnetic suspension densimeter due to the fluid-specific effect and its correction for use with gas mixtures containing oxygen
arXiv:2409.07140 · doi:10.1016/j.measurement.2019.107176
Abstract
Density measurements from single-sinker magnetic suspension densimeters need to be corrected to compensate for the magnetic effects of the measuring cell materials and the fluid on the coupling transmission system. While the magnetic effect of the densimeter materials can be easily determined, the fluid effect requires the calculation of an apparatus-specific constant, . In this work, the apparatus-specific constant of the single-sinker magnetic suspension densimeter at the University of Valladolid has been determined by using two alternative methods. The first method, which uses density data for the same fluid and conditions and different sinkers, yielded a value of = . The second method, obtained from measurements with pure oxygen, yielded a value of = . The second value is considered as more reliable, as the first method presents inherent limitations in this case.
References in corpus (1)
Cited by in corpus (5)
- Accurate experimental (, , ) data of the ( + ) binary system for the development of models for CCS processes
- Thermodynamic characterization of the ( + ) system significant for the hydrogen economy: Experimental (, , ) determination and equation-of-state modelling
- Thermodynamic characterization of the ( + ) binary system for the development of models for CCS processes: Accurate experimental (, , ) data and virial coefficients
- Thermodynamic () characterization of a reference high-calorific natural gas mixture when hydrogen is added up to 20 % (mol/mol)
- Impact of hydrogen addition, up to 20 % (mol/mol), on the thermodynamic (, , ) properties of a reference high-calorific natural gas mixture with significant ethane and propane content