Spatial resolution studies using point spread function extraction in optically read out Micromegas and GEM detectors

arXiv:2407.15491

Abstract

Optically read out gaseous detectors are used in track reconstruction and imaging applications requiring high granularity images. Among resolution-determining factors, the amplification stage plays a crucial role and optimisations of detector geometry are pursued to maximise spatial resolution. To compare MicroPattern Gaseous Detector (MPGD) technologies, focused low-energy X-ray beams at the SOLEIL synchrotron facility were used to record and extract point spread function widths with Micromegas and GEM detectors. Point spread function width of 108\,\microns for Micromegas and 127\,\microns for GEM foils were extracted. The scanning of the beam with different intensities, energies and across the detector active region can be used to quantify resolution-limiting factors and improve imaging detectors using MPGD amplification stages.

22 pages, 17 figures