Ultra-High precision Compton polarimetry at 2 GeV
arXiv:2402.16135 · doi:10.1103/PhysRevC.109.024323
Abstract
We report a high precision measurement of electron beam polarization using Compton polarimetry. The measurement was made in experimental Hall A at Jefferson Lab during the CREX experiment in 2020. A total uncertainty of dP/P=0.36% was achieved detecting the back-scattered photons from the Compton scattering process. This is the highest accuracy in a measurement of electron beam polarization using Compton scattering ever reported, surpassing the ground-breaking measurement from the SLAC Large Detector (SLD) Compton polarimeter. Such uncertainty reaches the level required for the future flagship measurements to be made by the MOLLER and SoLID experiments.