Polarization Leakage and the IXPE PSF
arXiv:2401.10231 · doi:10.3847/1538-4357/ad2065
Abstract
By measuring photoelectron tracks, the gas pixel detectors of the Imaging X-ray Polarimetry Explorer satellite provide estimates of the photon detection location and its electric vector position angle (EVPA). However, imperfections in reconstructing event positions blur the image and EVPA-position correlations result in artificial polarized halos around bright sources. We introduce a new model describing this "polarization leakage" and use it to recover the on-orbit telescope point-spread functions, useful for faint source detection and image reconstruction. These point spread functions are more accurate than previous approximations or ground-calibrated products ( and respectively for a bright -count source). We also define an algorithm for polarization leakage correction substantially more accurate than existing prescriptions (). These corrections depend on the reconstruction method, and we supply prescriptions for the mission-standard "Moments" methods as well as for "Neural Net" event reconstruction. Finally, we present a method to isolate leakage contributions to polarization observations of extended sources and show that an accurate PSF allows the extraction of sub-PSF-scale polarization patterns.
arXiv commentAccepted to ApJ. 9+3 pages. Revision corrected typos in Eq. 7 and B2
References in corpus (7)
- An efficient photoelectric X-ray Polarimeter for the study of Black Holes and Neutron Stars
- ixpeobssim: a Simulation and Analysis Framework for the Imaging X-ray Polarimetry Explorer
- Analyzing the Data from X-ray Polarimeters with Stokes Parameters
- Deep Ensemble Analysis for Imaging X-ray Polarimetry
- Towards Optimal Signal Extraction for Imaging X-ray Polarimetry
- Polarisation Leakage due to Errors in Track Reconstruction in Gas Pixel Detectors
- Improved Measurements of the IXPE Crab Polarization