Characterizing and correcting electron and hole trapping in germanium cross-strip detectors
arXiv:2306.08778 · doi:10.1016/j.nima.2023.168562
Abstract
We present measurements of electron and hole trapping in three COSI germanium cross-strip detectors. By characterizing the relative charge collection efficiency (CCE) as a function of interaction depth, we show that intrinsic trapping of both electrons and holes have significant effects on the spectroscopic performance of the detectors. We find that both the electron and hole trapping vary from detector to detector, demonstrating the need for empirical trapping measurements and corrections. Using our measurements of charge trapping, we develop a continuous depth-dependent second-order energy correction procedure. We show that applying this empirical trapping correction produces significant improvements to spectral resolution and to the accuracy of the energy reconstruction.
arXiv comment17 pages, 3 figures, 1 table, in press in NIM A
References in corpus (6)
- SciPy 1.0--Fundamental Algorithms for Scientific Computing in Python
- The Compton Spectrometer and Imager
- Calibrations of the Compton Spectrometer and Imager
- Impact of Charge Trapping on the Energy Resolution of Ge Detectors for Rare-Event Physics Searches
- Charge Trapping and Energy Performance of the MAJORANA DEMONSTRATOR
- Bayesian inference of high-purity germanium detector impurities based on capacitance measurements and machine-learning accelerated capacitance calculations