HyperSLICE: HyperBand optimized Spiral for Low-latency Interactive Cardiac Examination
arXiv:2302.02688 · doi:10.1002/mrm.29855
Abstract
PURPOSE: Interactive cardiac magnetic resonance imaging is used for fast scan planning and MR guided interventions. However, the requirement for real-time acquisition and near real-time visualization constrains the achievable spatio-temporal resolution. This study aims to improve interactive imaging resolution through optimization of undersampled spiral sampling and leveraging of deep learning for low-latency reconstruction (deep artifact suppression). METHODS: A variable density spiral trajectory was parametrized and optimized via HyperBand to provide the best candidate trajectory for rapid deep artifact suppression. Training data consisted of 692 breath-held CINEs. The developed interactive sequence was tested in simulations and prospectively in 13 subjects (10 for image evaluation, 2 during catheterization, 1 during exercise). In the prospective study, the optimized framework -HyperSLICE- was compared to conventional Cartesian real-time, and breath-hold CINE imaging in terms quantitative and qualitative image metrics. Statistical differences were tested using Friedman chi-squared tests with post-hoc Nemenyi test (p<0.05). RESULTS: In simulations the NRMSE, pSNR, SSIM and LAPE were all statistically significantly higher using optimized spiral compared to radial and uniform spiral sampling, particularly after scan plan changes (SSIM: 0.71 vs 0.45 and 0.43). Prospectively, HyperSLICE enabled a higher spatial and temporal resolution than conventional Cartesian real-time imaging. The pipeline was demonstrated in patients during catheter pull back showing sufficiently fast reconstruction for interactive imaging. CONCLUSION: HyperSLICE enables high spatial and temporal resolution interactive imaging. Optimizing the spiral sampling enabled better overall image quality and superior handling of image transitions compared to radial and uniform spiral trajectories.
References in corpus (12)
- SciPy 1.0--Fundamental Algorithms for Scientific Computing in Python
- Hyperband: A Novel Bandit-Based Approach to Hyperparameter Optimization
- Advancing machine learning for MR image reconstruction with an open competition: Overview of the 2019 fastMRI challenge
- Real-time Cardiovascular MR with Spatio-temporal Artifact Suppression using Deep Learning - Proof of Concept in Congenital Heart Disease
- J-MoDL: Joint Model-Based Deep Learning for Optimized Sampling and Reconstruction
- Spatio-Temporal Deep Learning-Based Undersampling Artefact Reduction for 2D Radial Cine MRI with Limited Data
- B-spline Parameterized Joint Optimization of Reconstruction and K-space Trajectories (BJORK) for Accelerated 2D MRI
- Accelerated Computing in Magnetic Resonance Imaging -- Real-Time Imaging Using Non-Linear Inverse Reconstruction
- Free-breathing and ungated cardiac cine using navigator-less spiral SToRM
- HyperSLICE: HyperBand optimized Spiral for Low-latency Interactive Cardiac Examination
- Extending LOUPE for K-space Under-sampling Pattern Optimization in Multi-coil MRI
- Learning Task-Oriented Flows to Mutually Guide Feature Alignment in Synthesized and Real Video Denoising