Effect of Diffusion on the Peak Value of Energy Loss Observed in a LArTPC
arXiv:2205.06745 · doi:10.1088/1748-0221/17/10/P10044
Abstract
Liquid Argon Time Projection Chamber (LArTPC) detectors observe ionization electrons to measure charged particle trajectories and energy. In a LArTPC, the long time (ms) between when the ionization is produced and when it is collected means that diffusion can smear the charge by an amount comparable to the spatial resolution of the detector, given by the spacing between charge sensing channels (mm). This smearing has an impact on the distribution of energy losses measured by each channel. In particular, the smearing increases the length of the charged particle trajectory observed by each channel, and therefore the most-probable-value (MPV) of particle energy loss recorded by that channel. We find, for example, that this effect shifts the MPV of a muon with an energy of 1 GeV by % for a 2 ms drift time and 4.7 mm wire spacing, as in the DUNE-FD LArTPC. This has implications for the energy-scale calibration and electron lifetime measurements in a LArTPC, which both use the MPV of the muon energy loss distribution as a "standard candle". The impact of diffusion on these calibrations is assessed.
References in corpus (11)
- Design and Construction of the MicroBooNE Detector
- A Proposal for a Three Detector Short-Baseline Neutrino Oscillation Program in the Fermilab Booster Neutrino Beam
- The ArgoNeuT Detector in the NuMI Low-Energy beam line at Fermilab
- First results on ProtoDUNE-SP liquid argon time projection chamber performance from a beam test at the CERN Neutrino Platform
- A study of electron recombination using highly ionizing particles in the ArgoNeuT Liquid Argon TPC
- Ionization Electron Signal Processing in Single Phase LArTPCs I. Algorithm Description and Quantitative Evaluation with MicroBooNE Simulation
- Calibration of the charge and energy loss per unit length of the MicroBooNE liquid argon time projection chamber using muons and protons
- Measurement of Longitudinal Electron Diffusion in Liquid Argon
- Measurement of Space Charge Effects in the MicroBooNE LArTPC Using Cosmic Muons
- The Liquid Argon In A Testbeam (LArIAT) Experiment
- Analysis of a Large Sample of Neutrino-Induced Muons with the ArgoNeuT Detector