A capacitive displacement system for studying the piezoelectric strain and its temperature variation
arXiv:2104.02211 · doi:10.1063/5.0040541
Abstract
A capacitive displacement system was constructed to measure the electric-field-induced piezoelectric strain in the simple form of either a bulk or thin film. The system can determine an AC displacement of 2 pm precisely by using a lock-in detection, and can measure the large displacement within a range of 25 m with a sub-nanometer resolution. The system can also be used to measure the variation in strain within a temperature range of 210 - 450 K, allowing the evaluation of the temperature coefficient of a piezoelectric constant and the studies on the effects of a phase transition on the piezoelectric response. Experimental results on quartz, PZT ceramics and thin films, and BaTiO confirm the capabilities of the developed system.
11 pages, 8 figures