Measurement of the distribution of Bi depositions on calibration sources for SuperNEMO
arXiv:2103.14429 · doi:10.1088/1748-0221/16/07/T07012
Abstract
The SuperNEMO experiment will search for neutrinoless double-beta decay (), and study the Standard-Model double-beta decay process (). The SuperNEMO technology can measure the energy of each of the electrons produced in a double-beta () decay, and can reconstruct the topology of their individual tracks. The study of the double-beta decay spectrum requires very accurate energy calibration to be carried out periodically. The SuperNEMO Demonstrator Module will be calibrated using 42 calibration sources, each consisting of a droplet of Bi within a frame assembly. The quality of these sources, which depends upon the entire Bi droplet being contained within the frame, is key for correctly calibrating SuperNEMO's energy response. In this paper, we present a novel method for precisely measuring the exact geometry of the deposition of Bi droplets within the frames, using Timepix pixel detectors. We studied 49 different sources and selected 42 high-quality sources with the most central source positioning.
16 pages, 12 figures, submitted to JINST, response to reviewer comments